Radwag MA 200.3Y Moisture Analyzer Manuale Utente Pagina 89

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 170
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 88
- 89 -
The display indicates initial mass of tested sample. In order to ensure stable
measurement conditions, it is recommended to close the drying chamber’s lid
before tarring the indication.
Press key to tare drying pan mass, and save its mass in moisture analyzer’s
memory.
Mass indication returns to ZERO and the following command is displayed.
If <START MASS CONTROL> option is enabled, at this moment the display
activates a bar graph indicating current sample mass in relation to saved Min
and Max limits for sample’s mass.
Load tested sample onto the drying pan. Select sample’s mass according to
previously carried out tests and characteristics of tested substance / material.
Vedere la pagina 88
1 2 ... 84 85 86 87 88 89 90 91 92 93 94 ... 169 170

Commenti su questo manuale

Nessun commento