An example of a report from drying chamber test:
--------- Drying chamber test ---------
Start time 2012.10.30 13:57:05
Balance model MA 3Y.NH
Balance ID 1352
Adjustment kit no.
Target temperature 120 °C
End temperature 119 °C
Measured temperature 121 °C
Permissible error +/- 3 °C
Status OK
---------------------------------------
The report is not stored in moisture analyzer memory. Exiting the report
immediately deletes it from the moisture analyzer volatile memory.
If a user aborts the testing process, the summary window will not be displayed,
and the software automatically returns to displaying the menu window
<Adjustment>.