28.4.15. Unlock the scale keyboard
Syntax: K0 CR LF
Possible answers: K0_OK CR LF – command in progress
28.4.16. Initiating of dosing/filling
Syntax: S1 CR LF
Possible answers:
S1_I CR LF
- command comprehended but cannot be executed
S1_OK CR LF
- command in progress
28.4.17. Stop of dosing/filling
Syntax: S0 CR LF
Possible answers:
S0_I CR LF
- command comprehended but cannot be executed
S0_OK CR LF
- command in progress
28.4.18. Set lower threshold
Syntax: DH_XXXXX CR LF, where: XXXXX – mass format
Possible answers:
DH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
28.4.19. Set upper threshold
Syntax: UH_XXXXX CR LF, where: XXXXX – mass format